Testing and quality assurance
Zhengxi equipment has been fully tested and verified, including:
AC/DC on-off and life testing
Reliability, flashover breakdown and comprehensive performance testing
High and low temperature, humidity and heat, drop and vibration testing
Ultrasonic non-destructive testing and plastic mechanics, electrical performance testing
Silicon steel sheet loss and chemical analysis and other tests to ensure superior product performance.
Main technical parameter
| Product Model | Input Voltage | Output Voltage | Size L X W X H(Mm) |
| ES710-4KVA | 220V | 220V | 218*290*185 |
| ES710-5KVA | 260*330*235 | ||
| ES710-6.3KVA | 260*330*235 | ||
| ES710-8KVA | 260*355*280 | ||
| ES710-10KVA | 410*440*350 | ||
| ES710-15KVA | 520*470*420 |
Scope of application
Medical isolation transformers are used to isolate the protected area of the medical area from the non-protected area circuit to avoid equipment damage or micro-electric shock risks caused by dangerous currents. Its safe and stable power supply ensures the reliable operation of medical equipment, and is particularly suitable for key medical places such as operating rooms and ICUs.
Product Features
Insulation grade: F/H grade (maximum working environment temperature 46°C)
Protection grade: IP00
Alarm mode: sound and light alarm, optional telephone alarm
Insulation measurement range: 0-999kΩ
Cooling method: fan cooling (applicable to above 40°C)
Operating temperature: -10°C to +60°C
Low noise operation: noise less than 35 dB
Compliance with standards: strictly in accordance with GB16895.24-2005 and GB19212.16 standards
This isolation transformer is tailored for medical places, providing solid guarantee for safe and reliable power supply, and is an ideal choice for medical power systems.
ES710 single-phase and three-phase isolation transformer for medical equipment.
Product analysis

② Digital Display
③ Metal Case
④ Security lock
⑤ Heat Sink